default

Request a Quote

Christmas-Card-2024

Dear valued customer,

As the Holiday Season approaches, we want to keep you informed about our upcoming closures to ensure a seamless experience for you and your laboratory.

Please note that our office operations and order processing department will be closed during these times:

Christmas Day Closure:
Wednesday, December 25, 2024

 

New Year's Day Closure:
Wednesday, January 1, 2025


Half-Days:
Tuesday, December 24, 2024
Tuesday, December 31, 2024

We will be available to take your calls and orders until noon.

To facilitate a smooth transition during these closures, we kindly request that you plan your orders, requests, and inquiries accordingly. If you have any questions or require further information, please don't hesitate to contact our customer support team at custservice@hardydiagnostics.com or call us toll-free at 800-266-2222.

We appreciate your cooperation and understanding during this holiday period. Thank you for choosing Hardy Diagnostics as your trusted partner. All of us at Hardy Diagnostics would like to express our gratitude for your continued support and reliance on our services! From our lab to yours, Happy Holidays!

NG-Test MCR-1, Research Use Only

Cat. no.

NGBMCRS23000

PACK SIZE:

20 per package

For Research Use Only


Rapid test for the detection of Colistin resistance via expression of mcr-1 gene on bacterial colony from culture. 

Not for in vitro diagnostic use

Performance Characteristics

Detection limit
The detection limit was determined using purified recombinant enzymes MCR detection threshold: 300pg/mL.

Validation on a reference strain bank
NG-Test MCR-1 was evaluated on 44 clinical strains at NCR Kremlin bicêtre-Paris (AMR French Reference Center), 117 clinical strains at ANSES-Lyon (National Health Security Agency) and 123 clinical strains at CNR-Clermont Ferrand. (Reference National Center).

A total of 284 Strains were evaluated.

Sensitivity: 100%   Confidence interval 95% : 96.3% to 100%
Specificity: 98.3%   Confidence interval 95% : 95.3% to 99.4%

Considering targeted MCR-1 enzyme, all results were correlated to the genotype of strains determined by PCR analysis.

NG-Test MCR-1 detects some MCR-2 variants.

More Information
Industry Clinical
Manufacturer NG Biotech
Pack Size 20 per package